Fix Transducers on Pipe
Using Mounting Belts
The transducers are mounted by using the metal ten-
sion belts. The length of the belts are related to the
maximum pipe size of the corresponding transducer (for
example, when using –F10 (1 1/4" to 16") size of belt
is suited for max. 16"). If using the same transducer for
smaller pipes, shorten the belts. The belts are fastened
by a tightener (Figure 37) which can be tightened by
using a screwdriver (Figure 38).
FIGURE 37:
Tightener for Metal Tension Belts
FIGURE 39:
V-mode or W-mode without spacer bar
FIGURE 40:
V-mode or W-mode with spacer bar
FIGURE 38:
Tighten metal belts with screwdriver
Z-mode Mounting of Ultrasonic
FIGURE 41:
V-Mode Mount with Fabric-Reinforced
Tensioning Tapes for Large Pipe Diameters