APPLIES TO FIXED UFM
21
FIGURE 35:
Lower side of ultrasonic transducer (touching
pipe wall)
Grease coupling
before mounting
Mounting in V-mode or W-mode
After setting up the parameters of the measuring point,
the flow transmitter displays the distance between
the transducers in inches (face-to-face, see Figure
36) and as number of grid holes for use of the spacer
bar (ultrasonic transducers type F10 and F21). Bar
index number 4, for example, shows three free grid
holes between the ultrasonic transducers. Install the
transducers on the spacer bar as shown in the figure.
Secure the transducers on the spacer bar using the
knurled screws .
FIGURE 36:
Using spacer bar – bar index holes
numbering, Example number of holes = 4
FIGURE 33: Z-mode installation with fabric-reinforced
tensioning tapes for large pipe diameters
ULTRASONIC TRANSDUCER
MOUNTING
The ultrasonic transducer (F10 and F21) is made of
plastic (PEEK) that has a beige color and is protected
by means of a metal sheath. The sheath has two
threaded bars on top that prepares the transducer for
use of spacer bar.
The ultrasonic transducer type F05 consists only of the
sensor sheath and the ultrasonic transducer.
FIGURE 34:
View of the back side of the transducer.
Some coupling grease (approximately the size of a
peanut) has to be applied where the part touches the
pipe wall after installation.
Ultrasonic Transducer
Cable
Sheath