Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 31 / 75
5.9.2
5.9.2
5.9.2
5.9.2
Fast Peak Value Recording
Fast Peak Value Recording
Fast Peak Value Recording
Fast Peak Value Recording
The fast recording of the maximum values of DC
fields is required for short magnetic impulses, as they
are generated in e.g. magnetizing systems.
In this operating mode you record peak values of the
magnetic field starting from a duration of 10µs. The
display shows the absolute highest amount.
Reset the value via the
NULL
button.
The automatic range selection cannot be used in this operating mode.
Note
Please note that in sensitive measuring ranges magnetic AC fields
caused by e.g. transformers or electric lines might lead to a measuring
value. Normally, these values would not be noticed during normal op-
eration, because disturbances with a net frequency of 50 or 60 hertz
are filtered out.
Operation via the external interface:
:PEAK:MODE FAST
5.9.2.1
5.9.2.1
5.9.2.1
5.9.2.1
Fast Peak Value Measurement Characteristics
Fast Peak Value Measurement Characteristics
Fast Peak Value Measurement Characteristics
Fast Peak Value Measurement Characteristics
Maximum ratings for the fast peak value measurement
Range
≤±1.0 %
≤±2.0 %
10 mT
≤70 Hz
≤100 Hz
100 mT
≤100 Hz
≤150 Hz
1 T
≤300 Hz
≤500 Hz
4.5 T
≤500 Hz (B <1.5T)
≤700 Hz