Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 30 / 75
5.9
5.9
5.9
5.9
Peak Value Measurement
Peak Value Measurement
Peak Value Measurement
Peak Value Measurement
The device has 2 different operating modes for measuring peak values. They differ in
speed, evaluation and resolution.
5.9.1
5.9.1
5.9.1
5.9.1
Normal Peak Value Recording
Normal Peak Value Recording
Normal Peak Value Recording
Normal Peak Value Recording
For the normal peak value recording of minimum and
maximum values the displayed measuring value is
analyzed continuously and the lowest or respectively
highest value are determined from it.
In this operating mode approx. 10 measurements per
second are carried out in high resolution. This enables
recordings of slowly changing field sizes. A typical application is the determination of
a maximum field strength value of a permanent magnet by manual positioning of the
measuring probe on the surface.
Reset the value via the
NULL
button.
The automatic range selection cannot be used in this operating mode.
The accuracy corresponds to the DC field measurement.
Operation via the external interface:
:PEAK:MODE SLOW