Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 16 / 75
3.2.4
3.2.4
3.2.4
3.2.4
External Static Magnetic Fields
External Static Magnetic Fields
External Static Magnetic Fields
External Static Magnetic Fields
Particularly in sensitive measuring ranges, an external static magnetic field, as e.g. the
Earth’s field, can already become clearly noticeable. These external magnetic fields lead
to a corruption of the measuring result.
To compensate external magnetic stray fields or asymmetries of the Hall probe, the de-
vice can be reset.
For this purpose please hold the measuring probe into a field-free area, e.g. a zero
Gauss chamber, or orient the measuring probe in a free field in east-west direction and
press the NULL
NULL
NULL
NULL button for approx. 3 seconds. The device then carries out a null balance
automatically.
The values are stored so that this balancing has to be carried out only in seldom cases.
If the magnetic field is too high during the automatic balancing, the correction is stopped
with an error message.