Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 29 / 75
Frequency Response Factor (Sinusoidal Field Pattern)
Frequency
Factor
2 kHz
1.00
5 kHz
0.98
7 kHz
0.95
10 kHz
0.90
Shape Factor (Sinusoidal Field Pattern)
Field Strength
B
eff
Factor
700 mT
1.00
1000 mT
1.01
1500 mT
1.02
2000 mT
1.03
The frequency response factor and shape factor are multiplied with the basic accu-
racy in dependence of the respective measuring parameters, and then added to the
basic accuracy.
Example:
Measuring
B
eff
= 1000 mT at a frequency of 5 kHz.
The basic accuracy is 0.5%. The frequency response factor is 0.9%. The shape factor
is 1.02.
The error from of these factors is 0.95 * 1.02 = 0.97 = -3%
For the measurement you have to calculate a total error of -3%±0.5%.