GMC-I Messtechnik GmbH
49
OUTPUT, OUTPUT? – Switching the Power Output On and Off
OUTPUT / LED
Function
The power output can be activated and deactivated with the
OUTPUT function.
Activation:
OUTPUT ON:
Current and voltage values of “0” are specified initially for a period
of approximately 2 ms
with activated output
for the transition from
the “highly resistive” condition. The output is then adjusted to the
selected voltage and current setpoints.
Deactivation:
OUTPUT OFF:
The power output is deactivated and rendered highly resistive
with the OUTPUT OFF command. Differentiation must be made
as to whether the internal dynamic sink is on or off (setting
command: SINK on / SINK OFF).
However, the output terminals are not electrically enabled.
OUTPUT OFF with SINK ON
The setpoints for voltage and current are set to 0 V and 0 A. The
sink is activated for approximately 300 ms. The sink discharges
the output capacitors as far as possible. The sink is then switched
off and the output becomes highly resistive as a result.
OUTPUT OFF with SINK OFF
The setpoints for voltage and current are set to 0 V and 0 A. The
power output is deactivated and becomes highly resistive as a
result.
The output capacitors are discharged via the connected load
only. Output voltage is correspondingly reduced.
Setting Command
Syntax:
OUTPUT txt
Parameter txt:
OFF/ON
Default setting
or after RESET (*RST): OFF
Query Command
Syntax:
OUTPUT?
Sample response string:
OUTPUT ON
Parameters List
Comments
If the output is switched off by a trigger signal in the
“T_MODE OUT” operating mode, i.e. in the off state, the signal
has higher priority.
An OUTPUT ON command is not executed, and bit 4 is set in
event register B (OUTE).
“Err 073” also appears briefly at the display as a warning in the
event of manual operation.
Additional functions which may influence the status of the output
include:
– OTP, overtemperature protection
– OVP, overvoltage protection
– OCP with activated “OCP ON” parameter
– SEQUENCE function
– T_MODE function
– POWER_ON
– *RCL
OVP, OVP? – Overvoltage Protection
Menu, LEDs
Edit Formulation
Function
The OVP function (
o
ver
v
oltage
p
rotection) specifies how the
power output will respond if output voltage is equal to or exceeds
the selected OVSET value.
The OVP function is an overriding protective function and is
independent of the voltage and current regulators.
The OVP function protects connected power consumers against
continuous overvoltage, although higher voltage is required
intermittently. The function also makes it possible to activate
another device configuration in case of overvoltage.
Output voltage is compared with the OVSET value from the
OVP-DAC by an autonomous comparator.
The ensuing reaction is shown in the following table, and
execution can be delayed by means of OV_DELAY.
Where OV_DELAY = 0, the power output is switched off directly
by the OVP comparator.
Activation of the OVP function is indicated at the front panel by
means of the “OVP ON” LED.
If OVP has caused shutdown, this is additionally indicated by the
“OVP” LED.
Setting Command
Syntax:
OVP txt
Parameter txt:
OFF/ON/R01 ... R12/15
Default setting
or after RESET (*RST): ON
Query Command
Syntax:
OVP?
Sample response string:
OVP OFF
Parameters List
OV_DELAY, OV_DELAY? – Overvoltage Protection Triggering Delay
Menu
Function
Desired response delay for the OVP function is set with OV_DE-
LAY. Delay time is specified in seconds.
If output voltage drops to below the OVSET value before OV_DE-
LAY expires, the shutdown sequence is interrupted and restarted
the next time the threshold is exceeded.
Where OV_DELAY = 0, the OVP comparator switches the power
output off directly as well.
Setting Command
Syntax:
OV_DELAY w
Value range:
0
w
65,535
Default setting
or after RESET (*RST):
0
Query Command
Syntax:
OV_DELAY?
Sample response string:
OV_DELAY XX.XXX
Param-
eter
Content Meaning
txt
OFF
Output is switched off,
OUTPUT LED off, control mode LEDs are off
ON
Output is switched on,
OUTPUT LED lights up, control mode LED lights up
Param-
eter
Content Meaning
txt
OFF
OVP function inactive
ON
OVP function activated:
The output is shut down if output voltage is equal to or exceeds
the specified OV_SET limit value for a duration of OV_DELAY.
Rxx
Instead of a shutdown, recall of a device configuration from
setup memories 01 through 12/15 can be activated with
R01 ... R12/15.