Technical Data
1)
T he voltage measuring ranges 100 mV .... 10 V can also be used for current measurements
as shown in the following table (e.g. reverse currents or leak currents). The accuracy of
these measuring ranges then corresponds to class 5.
Frequency range on
~
: 15 ... 100 Hz
Overload protection same as on the voltage measuring ranges.
2)
Measuring value = indicated value + constant ; 0 dB
0.775 V, i.e. 1 mW at 600
3)
protected by PTC thermistor; cooling time of thermistor before the start of next measure-
ment: grant at least 2 minutes.
Voltage
1)
Level
2)
Input
Resistance R
i
Overload
Protection
up to
3)
Const.
Measuring Span
100 mV
–20 dB
–40 ... –1 8 dB
10 M
/ /50 pF
600 V
300 mV
–10 dB
–30 ... – 8 dB
600 V
1 V
0 dB
–20 ... + 2 dB
600 V
3 V
+10 dB
–10 ... + 12 dB
600 V
10 V
+20 dB
0 ... + 22 dB
600 V
30 V
+30 dB
+10 ... + 32 dB
600 V
100 V
+40 dB
+20 ... + 42 dB
600 V
300 V
+50 dB
+30 ... + 52 dB
600 V
600 V
+60 dB
+40 ... + 62 dB
600 V
Range
Current at Full Scale Value
R
i
100 mV
10
nA
10 M
300 mV
30
nA
1
V
100
nA
3
V
300
nA
10
V
1
µA
30
V
3
µA
Current
Input Resistance R
i
Voltage Drop
U
Overload Protection up to
10 µA
10,0 k
100 mV
600 V
3)
100 µA
1,0 k
100 mV
600 V
3)
1 mA
100,0
100 mV
600 V
3)
10 mA
10,0
100 mV
600 V
3)
100 mA
1,4
140 mV
600 V
3)
1
A
480 m
0,480 V
600 V
3)
1 0
A
26 m
260 mV
600 V
4)