GMC-I Messtechnik GmbH
23
3)
Protected by G type melting fuse link FF 1.6A/600V AC in conjunction
with protective power diodes.
4)
Fuse protection FF 16A/600V AC; 10 A: max. 10 min.
Accuracy at reference conditions per IEC 60051/EN 60051
Class 1.5 for DC variables; class 2.5 for AC variables and class 1.5 for
resistance (inherent deviation with reference to the scale length of
69 mm) corresponding to maximum 10% inherent deviation from the
measured value in the range of the bolder drawn scale arc.
Reference Conditions
Ambient
temperature
+23
C
2 K
Relative humidity
40 ... 60%
Position of use
Device and scale horizontal
1
Frequency of
measured variable
45 ... 65 Hz
Waveshape of
measured variable
sinusoidal
Battery voltage
7.5 V
0.1 V
for other influencing
variables
per IEC 60051/EN 60051
The multimeter is calibrated in rms values. It operates with full-wave
rectification which valuates the arithmetic mean value.
Resistance
Range
Read-out Range
Mid-Scale
Value
(R
i
)
Open-
Circuit
Voltage
U
0
Short-
Circuit
Current
I
K
Overload
Protec-
tion up to
x
1
1
...
2 k
45,6
100 mV
2,2
mA
600 V
3)
x 10
10
... 20 k
456,0
100 mV
0,22 mA
600 V
3)
x 100
100
... 200 k
4,56 k
100 mV
22
µA
600 V
3)
k
x
1
1 k
...
2 M
45,6 k
1
V
22
µA
600 V
3)
k
x
10
10 k
... 20 M
456,0 k
1
V
2,2
µA
600 V
3)