48
GMC-I Messtechnik GmbH
Calibration Functions: V – Hz
–
Ω
– Temp – mA
– mA
6.5
Temperature Simulation [°C / °F]
The temperature simulator is capable of simulating
resistance
temperature detectors
(
RTD
) or
thermocouples
(TC)
with
specification of the external reference junction temperature.
Ð
Select the
Temp RTD
or
Temp TC
calibration function with the
rotary switch.
Ð
Switch the calibrator on by pressing the
ON / OFF | LIGHT
key.
The last selected temperature sensor is displayed.
Ð
Connect the device to be calibrated with the measurement
cables.
Ð
Set the temperature value:
Simulated resistance or simulated voltage is applied directly to the
output
!
Select the digit to be changed with the
keys, and change
the respectively selected digit with the
keys. As an
alternative, press and hold the
keys with the cursor at
any entry position until the higher value digits are scrolled
through as well.
Ð
The output can be activated with the
OUT | ENTER
key, or
deactivated [
Stdby
] once again.
Selecting Resistance Temperature Detector (RTD) or Thermocouple (TC)
for Fixed Value, Interval or Ramp Function
Ð
Switch to the fixed value, interval or ramp function menu by
pressing the
ZERO / SEL | ESC
key.
Ð
Select the [
select
sensor
] menu with the
keys.
Ð
Select the desired sensor with the
keys. Acknowledge
your selection with
OUT | ENTER
. The display is switched to the
window for temperature value entry, but the selected
sensor still appears in the auxiliary display.
The reference temperature is indicated in the auxiliary display.
Parameter Entries for Thermocouple Temperature Simulation
Select unit of measure °C or °F – SEt menu.
info
set
time
...
temp
intern
unit
set
°F
°C
Select internal reference temperature – SEt menu.
info
set
time
...
temp
extern
unit
set
temp
set
extern
intern
temp
intern
23
.
7
°C
Select and set external reference temperature – SEt menu.
info
set
time
...
temp
intern
unit
set
temp
set
intern
extern
extern
set
21
.
0
°C
temp
extern
22
.
4
°C
MEAS /CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
MEAS /CAL
SETUP
MEAS /CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
MEAS /CAL
SETUP
MEAS /CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
MEAS /CAL
SETUP
Содержание METRACAL MC
Страница 1: ...METRACAL MC Multimeter Calibrator Operating Instructions 3 349 566 03 7 2 15 ...
Страница 5: ...GMC I Messtechnik GmbH 5 Contact Persons ...
Страница 25: ...GMC I Messtechnik GmbH 25 Control Functions ...
Страница 65: ...GMC I Messtechnik GmbH 65 Device and Measuring Parameters ...
Страница 75: ...GMC I Messtechnik GmbH 75 Technical Data ...