GMC-I Messtechnik GmbH
41
V/Hz,
Ω
, Temperature,
and A/Hz Measurements
5.8.3 Direct Current Measurement with Current Clamp Sensor, mA DC
Voltage/Current Transformer Output
When a current clamp sensor is connected to the multimeter
( V input / mA input), all current displays appear with the
correct value in accordance with the selected transformation
ratio. The only prerequisite is that the current sensor is equipped
with at least one of the below listed transformation ratios, and that
the ratio has been previously selected in the following menu
(
Cl
ip
≠
0FF
) (see also section 7.4).
Current Clamp Setup Menu
1nFo
...
SET
rAtE
...
CLIP
1
/
10
/
100
/
1000
/
0ff
The maximum allowable operating voltage is equal to the nominal
voltage of the current transformer. When reading the measured
value, additional error resulting from the current clamp sensor must
also be taken into consideration. (default setting:
OFF
)
CL
IP
Transformer
Ratios
DMM Measuring Ranges
Clamp Types
300 mV
3 V
30 V
1:1
1mV/1mA
300,00 mA
3,0000 A
30,000 A
WZ12C
1:10
1mV/10mA
3,0000 A
30,000 A
300,00 A
WZ12B, Z201A
1:100
1mV/100mA
30,000 A
300,00 A
3000,0 A
Z202A
1:1000
1 mV/1 A
300,00 A
3000,0 A
(3 000,0 A)
Z202A, Z203A,
WZ12C
MEAS / CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
–
+
+
~
R
x
Current
003
.
50
A
1:1000
A
DC
R
i
= 1 M
Ω/
9 M
Ω
mA
Содержание METRACAL MC
Страница 1: ...METRACAL MC Multimeter Calibrator Operating Instructions 3 349 566 03 7 2 15 ...
Страница 5: ...GMC I Messtechnik GmbH 5 Contact Persons ...
Страница 25: ...GMC I Messtechnik GmbH 25 Control Functions ...
Страница 65: ...GMC I Messtechnik GmbH 65 Device and Measuring Parameters ...
Страница 75: ...GMC I Messtechnik GmbH 75 Technical Data ...