![Gossen MetraWatt METRACAL MC Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/gossen-metrawatt/metracal-mc/metracal-mc_operating-instructions-manual_2247241033.webp)
GMC-I Messtechnik GmbH
33
V/Hz,
Ω
, Temperature,
and A/Hz Measurements
5.5
Diode Testing
with Constant Current of 1 mA
Ð
Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
Ð
Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Refer to section 5.2.1 regarding testing for the absence of volt-
age with the help of the direct voltage measurement.
Ð
Set the rotary switch to
.
Ð
Connect the device under test as shown.
Attention!
!
Observe open-circuit voltage of 7 V during diode testing.
Circuits must be laid out accordingly.
Forward Direction and Short-Circuit
The instrument displays forward voltage in volts (display:
4 places). As long as voltage drop does not exceed the maximum
display value of 6 V, several series connected components or
reference diodes can be tested with a small reference voltage and
Z-diodes.
Reverse Direction and Interruption
The measuring instrument indicates overload
0L
.
Note
Resistors and semiconductor paths connected in parallel to
the diode distort measurement results!
–
+
+
Forward Direction
Reverse Direction
0 V !
0
.
654
0
.
L
to 6 V
Measuring range:
+
–
!
1
.
8569
V
DC
Содержание METRACAL MC
Страница 1: ...METRACAL MC Multimeter Calibrator Operating Instructions 3 349 566 03 7 2 15 ...
Страница 5: ...GMC I Messtechnik GmbH 5 Contact Persons ...
Страница 25: ...GMC I Messtechnik GmbH 25 Control Functions ...
Страница 65: ...GMC I Messtechnik GmbH 65 Device and Measuring Parameters ...
Страница 75: ...GMC I Messtechnik GmbH 75 Technical Data ...