22
Gossen Metrawatt GmbH
SECULIFE SP BASE
5. Touch
-
to decrease the factory-programmed R-curve.
The current adjustment value is displayed. (The current adjustment value does
not have a direct correlation to the displayed SpO
2
value on the DUT.)
6. Touch
Exit
to return to the main screen without changes.
7. Touch
Save
to save the new settings.
The operating screen displays the SpO
2
setting in red letters on the main screen
and with a red highlighted bar on the preset screen.
Note!
There are no absolute standards for pulse oximetry, and each manufacturer
has defined its own proprietary R-curve. For this reason, some oximeter
readings may be outside of an acceptable range.
8.5
Displaying Instrument Information
1. Touch
Setup
.
2. Select
Instrument Information
.
The current version of the software and any other applicable information is dis-
played.
8.6
Updating Firmware
Erases the firmware from the instrument in order to update to a newer version of
the firmware.
1. Touch
Setup
.
2. Select
Update Firmware
.
A warning message is displayed: „Warning! Pressing OK will erase current pro-
gram. New program must be loaded by the user.“
3. Touch
OK
to confirm.
The current firmware is erased from the instrument.
4. Touch
Back
to return to the setup menu without changes.