20
Gossen Metrawatt GmbH
SECULIFE SP BASE
8.3.3
Selecting preset Parameters
1. Touch the mode selection button to change to
PRESET MODE
.
A list of preset combinations of SpO
2
%, heart rate, and perfusion index is dis-
played.
2. Touch the upper scroll button to select the desired preset combination.
The selected combination is highlighted.
3. Touch the lower scroll button to select the desired monitor.
4. Touch the mode selection button to turn back to
MANUAL MODE
.
The selected combination is set.
8.4
Changing Setup Configuration
8.4.1
Changing Display Brightness
1. Touch
Setup
.
2. Select
Display Brightness
.
3. Slide the control bar with your finger to adjust the display brightness in a range
between 5 % to 100 % in 5 % increments.
4. Touch
Back
to save the new settings.
8.4.2
Adjusting the Volume of System Sounds
1. Touch
Setup
.
2. Select
Sounds
.
3. Slide the control bar with your finger to adjust the volume of all system sounds
in a range between 5 % to 100 % in 5 % increments.
4. Set the function switch to
off
to turn the sound off.
5. Set the function switch to
on
to turn the sound on again.
6. Touch
Back
to save the new settings.
8.4.3
Setting Auto Off Timer
Adjusts the timeout that will turn the instrument off after a period of inactivity.
The instrument is running on battery power.
1. Touch
Setup
.
2. Select
Auto Off Timer
.
3. Slide the control bar with your finger to adjust the timeout in a range between 5
to 100 minutes in 5 minute increments.
4. Set the function switch to
off
to disable the timeout.
5. Set the function switch to
on
to enable the timeout again.