METRACABLE TDR PRO
Gossen Metrawatt GmbH
6.12
Instrument Updates
Firmware and FPGA updates are made available for the instrument.
Firmware Update
In order to update the instrument’s firmware, the new firmware must first be downloaded
and transferred to the instrument. Updating is then started at the instrument.
The instrument and the PC/METRACABLE MANAGER are connected to each other
20.
1. Select
Configuration
from the menu.
The
Configuration
dialog appears.
2. View the
Firmware Version
in the
Update
panel and make a note of it.
3. Determine the newest firmware version by visiting the Gossen Metrawatt GmbH website
at:
http://www.gossenmetrawatt.com
4. Compare the two versions.
–
If they’re identical, the instrument is up-to-date and no action is required.
–
If the firmware on the website is newer, download it and proceed as follows.
5. Click the
Update Firmware
button in the
Configuration
dialog in the
Update
panel in the
METRACABLE MANAGER program.
The
Select Update File
dialog appears.
6. Select the downloaded firmware file.
The file is transferred to the instrument.
7. Exit the
Data Transfer
menu item at the instrument.
The Bluetooth
®
module is deactivated. The main menu appears.
8. Select
Settings
from the menu with the help of the scroll keys.
9. Acknowledge by pressing the
SEL
key.
The
Settings
menu is displayed.
10. Select
Program Update
from the menu with the help of the scroll keys.
11. Acknowledge by pressing the
SEL
key.
The update is executed.
Update Active
appears at the display during the update process.
A message indicating successful completion then appears.
The instrument’s firmware has been updated.
FPGA Update
In order to update the instrument’s FPGA, the new FPGA must first be downloaded and
transferred to the instrument. Updating is then started at the instrument.
The instrument and the PC/METRACABLE MANAGER are connected to each other
20.
1. Select
Configuration
from the menu.
The
Configuration
dialog appears.
2. View the
FPGA Version
in the
Update
panel and make a note of it.
3. Determine the newest FPGA version by visiting the Gossen Metrawatt GmbH website at: