METRACABLE TDR PRO
Gossen Metrawatt GmbH
1. Open the TDR menu by pressing and holding the
SEL
key.
The TDR menu appears.
2. Navigate to the
Freeze
submenu with the help of the
and
scroll keys.
3. Acknowledge by pressing the
SEL
key.
The
Freeze
menu appears.
4. Select
On
with the
and
scroll keys.
5. Acknowledge by pressing the
SEL
key.
The
Freeze
menu appears.
6. Return to the measurement function by pressing the
ESC
key.
The measurement is displayed and is frozen.
Reference Function
Two measurements can be compared at a glance with the help of the reference function.
The first measurement is saved as a reference curve in the background of the second mea-
surement.
Proceed as follows in order to save a TDR trace in the background during a measurement:
1. Perform the first measurement
2. Open the TDR menu during the measurement by pressing and holding the
SEL
key.
The TDR menu appears.
3. Navigate to the
Reference
submenu with the help of the
and
scroll keys.
4. Acknowledge by pressing the
SEL
key.
The
Reference
menu appears.
5. Select
On
with the
and
scroll keys.
6. Acknowledge by pressing the
SEL
key.
The
Reference
menu appears.
7. Return to the measurement function by pressing the
ESC
key.
The measurement is displayed in the background.
8. Perform the second measurement.
Both measurements are displayed and can be compared with each other.
5.5
Pinpointing and Identifying Faults
Please observe the following notes in the specific case of pinpointing and identifying faults.
Fault Finding Recommendation
When conducting fault finding at a cable, measurement should be performed from both
cable ends – in particular in the case of faults in open circuits in which the actual end of the
cable is not visible.
You can use the reference function in order to compare the two measurements with each
other
When measuring from both ends, the combined results should correspond to the expected
cable length. Even if the actual cable end is indicated at the display, the reflections down-
stream from the fault may be too uncertain for a conclusive analysis. In this case, measure-
ment from both ends provides a clearer picture with increased accuracy.