GOSSEN METRAWATT GMBH
23
Figure 4.5
LCD Test
•
Internal Memory Test
Key Sequence:
InFO
tiME
tESt
diSP
rAM
no
YES
buSY
Settings:
no = no memory test is performed
YES = the memory test is started without any additional security queries
After acknowledging
YES
with the
ENTER
key, the adapter’s internal memory is tested. Testing
takes about 5 minutes.
All
data stored to memory are deleted during the memory test.
buSY
appears at the display during testing.
PASS
appears at the display when the test has been completed. The adapter is then returned
to its initial display by pressing the
ENTER
key.
Caution:
Make sure that all recorded measurement values which need to be saved have
been uploaded and archived at a PC
before
starting the memory test.
!