GOSSEN METRAWATT GMBH
5
3.4
Configuring Multimeter Parameters
– for all METRA
Hit
2x with SI232-II or BD232
3.5
Receiving Live Measurement Data from the DMM
– via adapter, but without saving to memory in the adapter
SI and BD adapters can be mixed when using any METRA
Hit
2x. However, only SI adapters
may be utilized if METRA
Hit
1x/2x are used in combination. If more than 4 adapters are
used, set the baud rate to 19.2 kBd, or to 9.6 kBd for fewer than 4 adapters.
SI232-II
Adr: 1
Bd-PC: 9600
Bd-MM: 8192
Mode: pc
METRA
Hit
2x
only
Adr: 1
Adapter: SI232
* store/online
rate:
send off
Rotary Switch:
set to measuring
function
BD232
Adr: 2
Bd-PC 9600
fixed
METRA
Hit
2x
only
Adr: 2
Adapter: BD232
rate:
send off
Rotary Switch:
set to measuring
function
PC
METRAwin 10
Set/Device type..., Set/Channels ...
*
online
only with
METRA
Hit
28S or 29S,
otherwise select
store
.
SI232/-II
Adr: 1
Bd-PC: 9600
Bd-MM: 8192
Mode: on(line)
METRA
Hit
1x, 2x
Adr: 1
Adapt.: SI232
* store/online
rate:
send on
Rotary Switch:
set to measuring
function
max. 4 at once
SI232/-II
Adr: 2
Bd-PC: 9600
Bd-MM: 8192
Mode: on(line)
METRA
Hit
1x, 2x
Adr: 2
Adapt.: SI232
* store/online
rate:
send on
Rotary Switch:
set to measuring
function
SI232-II
Adr: 1
Bd-PC: 9600
Bd-MM: 8192
Mode: pc
METRA
Hit
2x only
Adr: 1
Adapt.: SI232
* store/online
rate:
send off
Rotary Switch:
set to measuring
function
METRAwin 10
Set/Device type..., Set/Channels ..., File/Start measurement
BD232/ RS232
Adr: 1
Bd-PC 9600
fixed
METRA
Hit
2x only
Adr: 1
Adapter: BD/RS
rate:
send off
Rotary Switch:
set to measuring
function
*
online
only with METRA
Hit
28S or 29S, otherwise set to
store
.
PC
PC
PC