Information About the Sensor
Each configured ATOS sensor, in scan direction, has defined 3D areas
within which a measuring object can be scanned. In the following,
such a 3D area is called "measuring volume" (MV).
The measuring volume determines the distance between sensor and
measuring object and the set of lenses to be used.
In practice, depending on the measuring task, different measuring
volumes might be required.
Ideally, you have camera and projector lenses that are preadjusted to
your measuring volume. In this case, you need to perform no or just
very few steps for adjusting the sensor.
A complete sensor setup is only required if the corresponding set of
lenses has never been adjusted before or is decalibrated (wrong ad-
justments).
For a complete sensor adjustment, you need to adjust the angle rela-
tions of the cameras and set the focus and aperture of the lenses.
Then, the complete system is calibrated by means of calibration
panels or calibration crosses. If the measuring volume is adjusted
successfully by calibration, you may start a measuring project.
The sensor adjustment is supported by the software. The software de-
termines all necessary parameter settings.
In order to achieve large, medium and small measuring volumes, the
sensors can be adapted to the camera positions
SO
(Small objects)
for small measuring volumes,
300
for medium measuring volumes and
500
for large measuring volumes, see table 4.1, with the help of ex-
changeable camera supports.