I S A – M e n u O p e r a t i o n
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 23 / 76
www.go-sys.de
Opens the input menu of the intensity (number of flashes per single measurement).
Switches to the input menu of the measurement path length of the sensor head in mm.
Only visible if the path length is used in an AMS formula (query command
ISA.Pathlength
).
The input is performed in the AMS configuration window (see
9.2.1 The Configuration
Window
).
Opens the display of the current absorbance spectrum.
Opens the display of the current raw spectrum.
Switches back to the Spectrometer menu.
Service menu II
Opens an input menu for a probe number
∗
,
commas are saved as a point.
The last recorded spectrum is saved as sample spectrum (reference
spectrum) under the entered sample number.
Starts a measurement (spectrum capture).
Saves the last captured spectrum as a sample spectrum (reference spectrum) under the
entered probe number (see above).
Switches to the Spectrometer menu.
∗
Also called probe name or sample number. Character set: Standard ASCII
The probe number is required for an application specific calibration in connection with a multi-parameter calibration, and is
stored in the JCAMP-DX format together with the spectral data.
see
Manual ISA and Process Spectrometer Commissioning - Maintenance - Service
there
4.3 Application-Specific Calibration
Intensity spec
Path length
Abs spectrum
Raw spectrum
<–
Set sample no.
Start measuring
Save sample
<–