I S A – A M S O p e r a t i o n
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 27 / 76
www.go-sys.de
Unit
Unit of the output value
More than 5 characters can’t be displayed at the BlueBox display.
Digits before
Factory preset, 5 pre-decimal places are displayed.
Digits after
Factory preset, 0 decimal places are displayed. Because integer counts are measured
here, there are no decimal places.
Interval
Measurement interval = time period between the end of a measurement and the end
of the next measurement, lowest value is 3.
1
The higher the interval is set, the fewer spectra are stored, which shortens the
download times and saves storage space. One absorbance spectrum needs 1540 Byte
(1548 with GPS data). GO Systemelektronik recommends a minimum interval of 60,
otherwise the lifetime of the xenon flash lamp will be shortened.
Min. Value
2
Lowest expected measuring value, factory preset 0
Max. Value
1
Highest expected measuring value, factory preset 30000
Average
The number of single measurements from which the arithmetic mean is derived. The
arithmetic mean values of the respective spectral values of the single measurements
result the measured spectrum.
Wait until the end of a measurement before making an input.
Opens a menu where you can define how the measurement values and states of actua-
tors are stored in the database. The determination has no effect on spectra itself and
affects only the MVR
3
of the spectra.
Transfers the formula from the input field to the BlueBox.
The button is inoperable here.
Opens a window to save the formula on the PC.
The button is inoperable here.
Opens a window for to load a saved formula from the PC.
The button is inoperable here.
Opens a window for to print these sensor-settings.
Opens a list of the current variables with their current values.
Closes the sensor setup window.
1
The measurement interval is calculated in advance from the set interval time and the configuration parameters. This results in
a low time drift of the recording times of the measured values.
2
If there is a overrun or underrun of the range, the measurement value is displayed and stored for real sensors. In the case of
virtual sensors, the overrun or underrun is marked by a "<" or ">", then the entered minimum or the entered maximum meas-
urement value is stored. In addition, the sensor icon is displayed as a warning sign in the AMS start window.
3
MVR
= Maximum digital Value of a Raw spectrum