I S A – M e n u O p e r a t i o n
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 18 / 76
www.go-sys.de
Switches to the diagram display. see
8.1.2.2 Diagram View
Switches to the sensor information. see
8.1.3 Sensor Information
Switches to the Main menu.
Spectrometer menu II
Switches to the input of the sensor name.
max. 20 signs
Is queried by other BlueBox programs
Switches to the input of the unit of the output value.
More than 5 characters can’t be displayed at the BlueBox display.
Switches to the input of a name for the measured parameter.
max. 20 signs
Switches to the input of the measurement interval = time period between the end of a meas-
urement and the end of the next measurement, lowest value is 3.
∗
GO Systemelektronik recommends a minimum interval of 60, otherwise the lifetime of the
xenon flash lamp will be shortened.
see
8.1.1 Interval and Average
Switches to the input of the average: The number of single measurements from which the
arithmetic mean is derived. see
8.1.1 Interval and Average
Switches to the input of the measurement resolution.
Input 1 corresponds to decimal places = 0 in the sensor setup window,
Input 0.1 corresponds to decimal places = 1 in the sensor setup window, etc.
Switches to the Main menu.
Spectrometer menu III
Measurement range lower limit, factory preset 0
Measurement range upper limit, factory preset 30000
Switches to the Main menu.
Wait until the end of a measurement before making an input.
∗
The measurement interval is calculated in advance from the set interval time and the configuration parameters. This results in
a low time drift of the recording times of the measured values.
Diagram
Info
<–
Name
Unit
Parameter
Interval
Average
Resolution
<–
Min. limit
Max. limit
<–