- There is no water flow in the
tank when the patient gets
closer (optical sensor).
- Damaged-sensor.
- Distance between the pa-
tient to the upper sensor
larger than 300mm.
- Dirt on the lens sensor.
- Chair fuse is burnt
- Request assistance by GNA-
TUS Technician.
- Get closer to the sensor
(less than 300 mm).
- Clean the sensor lens.
- Turn off the power supply
to the chair and request
asistance by the Technician.
Curring Light
-Equipment’s not working.
-Power cut.
-Chair’s fuse burned.
-Check power supply.
-Turn off the chair from
mains power and request a
Technician presence.
-Equipment is not polymeri-
zing resins.
-Resin is not appropriate
for LED’s photopolymerizer
wave length range.
-Resin residues in light cable.
-Get the indicated resin for
the photopolymerizer’s wave
length range, one with con-
tains photoinitiators based
on camphorquinone.
- Clean the light cable.
Bicarbonate Jet
- Refer to Owner's Manual of Jet Hand (available for viewing
and downloading via www.gnatus.com.br/manuais)
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
Problems
Probable cause
Solution