Master Valve*
The master valve is a safety device that aims to block / release the entry of water to the
dental set. It is of utmost importance to have interrupted the water supply to the dental set
in the end of the working day, which can be done through the key (ON/OFF - 04).
How to provision the reservoirs
Water - Syringe/Handpieces
Remove the reservoir (B) uncoiling it on
clockwise and make the replacement of water. After
the replacement put it back coiling on anticlockwise.
Always use filtered water or aseptic products.
Bio-System*
Remove the reservoir (A) uncoiling it on clockwise
and make the replacement. Use a chlorinated water
solution 1:500
Preparing the solution:
From a solution of hypochlorite of sodium at 1%,
a solution of chlorine at 500 p.p.m. is prepared.
How to prepare the solution: Take 25ml of
hypochlorite of sodium at 1% and dilute it in 500 ml
of water (1 to 20). Such solution should be prepared
daily.
IMPORTANT: Follow this proportion strictly to avoid damages in the equipment and to
have an efficient result in the disinfection.
A
B
Bio-System*
Remove hanpieces from terminals. Take terminals to
bowl or water unit’s sink.
Open the terminal’s spray valves completelly.
Press the Bio-system key for some seconds, to disinfect
the equipment’s components internally with disinfectant.
Then, press the command pedal for some seconds to
rinse, in order to eliminate the disinfectant residues that
could have remained.
Important:
Repeat this procedure before working day
and after each patient.
EQUIPMENT OPERATION
Bicarbonate Jet "Jet Hand"
For further information, please see the Jet Hand manual which comes with the product.
* Optional