- Remove two hoses
connected to reservoir (3)
and (4).
- Release the tank locks
cover (5).
- Lift the cover and
remove the filter by moving
it sideways (6) e (7).
- Remove the tank pulling it up (8).
- Wash the filter and the reservoir in current water.
- Clean the electrodes (9).
- Fit the filter and place the tube in the tank surrounding the electrodes (10).
Note: Do not miss this tube, because it avoids the foam electrodes.
- Making the inverse procedure, mount the reservoir and cover the Biovac Sec.