18
CONSERVAÇÃO E MANUTENÇÃO CORRETIVA E PRE-
VENTIVA
Note: Use gloves and other
systems of protection,
during the disinfection.
Attention: Do not use any
disinfectant spray, as the vapor
may be inflammable, or it may
cause injury.
The equipment should be calibrated routinely, as per the legislation in force in the country.
But never with a period exceeding 3 years. In order to protect your equipment, seek Gnatus
technical assistance for periodic revisions of preventive maintenance.
Preventive Maintenance
Corrective Maintenance
If the equipment has any abnormality, check if the problem is related to any of the items
listed in the item Unforeseen Events (situation, cause and solution).
If it is not possible to solve the problem, turn off the equipment, and request Gnatus
technical assistance.
Problem
Probable cause
Solution
- There is no suction,
when the ejector is taken
from the holder.
- Lever of adjustment of
emptying of the sucker is
shut.
- Circuit breaker in the “off”
position.
- Unplugged cable.
- Contact the electrodes
sealed by dirt or foam in the
reservoir.
- Proceed the adjustment of
the emptying through the
lever acoording to need.
- Turn on the circuit bre
-
aker.
- Plug the cable.
- Contact the electrodes
sealed by dirt or foam in the
reservoir.
Upon coming across any problem in operation, follow the instructions below to
check and repair the problem, and/or get in touch with your representative.
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS