English
OPERATION OF EQUIPMENT
The time to produce 4 liters of distilled water is
approximately 4 hours. The distiller will automatically
turn off when all the water is evaporated from the
chamber.
After a complete distillation cycle, do not open the condensation cover immediately.
Wait for the cover cools down for at least 20 minutes.
When using the distiller for the first time, discard the first cycle of distilled water,
so that the internal components are completely sterilized by this first cycle.
Avoid connecting the apparatus without water in the camera.
Activated coal filter installation
E
E - The distiller will automatically turn on in the first
use, when this does not occur use the power button
switch "SPSP".
A - Remove the nozzle guide located in the condensation cover ;
B - Prepare the activated coal filter to replace the nozzle guide, put 2 to 3 glass of water in
the filter before each cycle of distillation. After the filter is dampened fit it in the condensation
cover (C1).
Only will be necessary to use the activated coal filter if water is indicated for human
consumption
A
B
c1
29