English
Upon coming across any problem in operation, follow the instructions below to check
and repair the problem, and/or get in touch with your representative.
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
-Completely out of order
-P l u g disconnected
-Lack of Power supply.
-Condensation cover’s plug
disconnected to the device.
-"SPSP" key not activated.
-Connect the plug.
-Wait for normalization.
-Connect the plug.
-Press the SPSP power but
-
ton again.
-After a distillation cycle, a
low production of distilled
water was obtained.
-Low level of water in the
vat.
- Condensation cover is deta
-
ched to the evaporation
chamber.
-Regulate the water level.
-Couple the condensation
cover.
-Steam leaking.
-Condensation cover is det
ched to the evaporation
chamber.
-Damaged Seal ring.
-Damaged Cover set: Con
-
densation coil or mini fan.
-Couple the condensation
cover.
-Contact the Gnatus techni
-
cal assistance.
- Contact the Gnatus techni
-
cal assistance.
Problem
Probable cause
Solution
If the equipment has any abnormality, check if the problem is related to any of the items
listed in the item Unforeseen Events (situation, cause and solution).
If it is not possible to solve the problem, turn off the equipment, and request Gnatus
technical assistance.
Corrective Maintenance
In order to protect your equipment, seek Gnatus technical assistance for periodic revisions
of preventive maintenance.
Preventive Maintenance
CORRECTIVE AND PREVENTIVE MAINTENANCE AND
PRESERVATION
35