55
M3: START/STOP
Micromotor and apex locator
The micromotor starts to spin in the
beginning of the apex measurement (START/
STOP). When it reaches the pre-determined
position of the apex, the micromotor stops
to spin.
IN MODES M3, M4 AND M5 IT IS NOT NECESSARY THE USE OF THE
PEDAL TO BE USED WITH THE MICROMOTOR.
M5: SLOW/DOWN
Micromotor e apex locator
The micromotor starts to spin at the
beginning of the apex measurement (START/
STOP). When it reaches the apex pre-
determined position, the, the micromotor
retrogresses slowly (SLOW/DOWN).
M4: AUTO2
Micromotor and apex locator
The micromotor starts to spin at the
beginning of the apex measurement
(START/STOP). When it reaches the apex
pre-determined position, the micromotor
retrogresses automatically (AUTO2).
IN CASE THE MEASUREMENT CABLE IS NOT CONNECTED TO THE
APEX LOCATOR CONNECTOR THE ELECTRIC MICROMOTOR WILL
ONLY FUNCTION IN MODE M1.
EQUIPMENT OPERATION