A-1
TYPICAL OPERATING
PARAMETERS
Typical operating parameters are as follows:
Gas
Range
Resolution Zero Accuracy
Range
Stability
ppm
0 to 1000 ppm 5 ppm +/- 30 ppm 5% +/- 25 ppm
LEL
0 to 10%
0.1% +/- 0.5%
10 to 100% 1%
N/A 2% +/- 1% LEL
Volume 0 to 100%
1%
+/- 2% 1% +/- 1% Gas
Gas
Carbon
0 to 1000 ppm 1 ppm CO +/- 15 ppm +/-15 ppm + 5%
Monoxide
reading
EXT
0 to 1000 ppm 1 ppm 5%
Typically 20%
1000 to 9999 ppm 10 ppm
Notes:
All the values above are at normal temperature and pressure.
Humidity is between 0% and 95% RH (non-condensing).
Pressure changes at the inlet and exhaust are minimised as they
may cause transient changes in reading.