FIRST RESPONDER - USER HANDBOOK
5-8
Probe Handle Assembly - Part No. 13561
Fitting of an In-line filter assembly to the probe handle, as
shown in Figure 5-8, will minimise the danger of water
and / or dust ingress.
.
Fig. 5-8 Probe Handle and In-line Filter Assembly
Caution: The instrument should never be switched
on without suitable filters installed.
If a blockage occurs the SAMPLE FAULT indicator is
displayed. Check the sample line and filter assembly for
blockage. Press Switch One (
) to clear the SAMPLE
FAULT message. Replace the filter if the message does
not clear.