41
REV.01 del 06.12.2017
NOTES ON THE USE OF MICROCURRENT PROGRAMS
This paragraph refers to the use of microcurrent programs.
The microcurrent programs differ from normal TENS and EMS programs as follows:
- While conventional electrostimulation (e.g. TENS) uses current in the
milliamperes range, microcurrent electrostimulation uses currents in the
microampere range that are imperceptible by humans. During Microcurrents
programs,
it is normal that the user does not discern any stimulation
.
- When running a Microcurrents program,
use exclusively the special gray
cables connected to the outlets of channels 1 and 3
. If the cables are not
connected or are of the wrong type, it will not be possible to start the program.
Check the connections and the cables.
-
The Microcurrents programs have prefixed intensity levels
, therefore it is not
necessary to set them. When a Microcurrent program is activated, the
electrostimulator automatically brings the intensity to the correct level. This value
should not be altered during the execution of the program.
- The Microcurrents programs cannot be run in the "2+2 mode" with multiple
treatments. If one tries to select a Microcurrents program in "2+2 mode", the
electrostimulator will emit an error tone.
If, according to your therapist, you wish to modify the treatment protocol altering
the intensity, press and hold the UP and DOWN button for 3 seconds.