31
REV.01 del 06.12.2017
The device selects automatically a value of intensity equal to the double quantity of
the reobase. The amplitude of the impulse is set up automatically to the minimal
value. The operator must gradually increase the value of the amplitude of the
impulse. When the first motor response occurs visually, the amplitude of the impulse
must be confirmed with OK by the therapist.
Slope calculation
After storing the chronaxy value, you have to set the slope value of the impule by
modifying in this way the impulse shape from rectangular (default) to trapezoidal-
triangular. Use the P+ and P- buttons to modify the value of the slope and confirm
with OK. Continue by increasing the value intensity up to reach the stimulation. The
value of slope and, thereby, the value of intensity can be modified again in the
setting screen.
NOTE: The programs created in the diagnostics menu are saved in the
PROGRAMMING area where they can be executed.
Results
Chronaxy inferior to 1 millisecond: the muscle is normally innervated.
Reobase 20mA
2x Reobase 40mA
Chronaxy 1,10 mS
Name
Rate Ramp up: Intensity
90° 40 mA
Inserire pendenza?
Sì
Diagnostics 1-15
Enter the slope
Yes
Phase 1