PART THREE - OPERATION
SECTION 3 - ANALYZER CONFIGURATION
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
Rev. 6-1100
40
Select
TEMP ELEMENT
Type
☞
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.
Create the desired notation on the second line:
A.
Starting with extreme left character position, use
×
×
and
Ø
Ø
keys
to select the desired first character.
B.
Press
Ö
Ö
key
once to select the next character, and
use
×
×
and
Ø
Ø
keys
to select its desired character.
C. Repeat procedure until desired notation is displayed.
3.
Press
ENTER key
to enter the displayed notation.
When measuring pH, configure the analyzer for automatic
temperature compensation by defining the temperature
element, or fixed MANUAL temperature compensation.
When using MANUAL, determine and enter a specific tem-
perature.
When using the analyzer to measure ORP, this
function only defines the element used for temperature
measurement. There is no temperature compensation
for ORP measurement, which does not require it.
NOTE:
When a temperature element type has been se-
lected but the element is not connected to the
analyzer, a “WARNING: CHECK STATUS” mes-
sage will appear. To prevent or clears this message,
connect the element or select “MANUAL.”
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view the
four choices:
•
NTC300:
Configures analyzer for use with NTC 300
ohm thermistor temperature element (used in GLI
5-wire Differential pH and ORP sensors).