PART THREE - OPERATION
SECTION 3 - ANALYZER CONFIGURATION
Rev. 6-1100
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
35
S
ECTION
3
☞
3.1 Selecting LANGUAGE
to Operate Analyzer
☞
3.2 Configuring
Sensor Characteristics
SELECT SENSOR
Type
NOTE:
When the passcode feature is enabled (Section
3.6), you must successfully enter the passcode
before attempting to enter a configuration setting.
The analyzer is equipped to display operating screens in
various languages such as English, French (Français),
German (Deutsche), Spanish (Español), and others. The
analyzer is factory-set for English. To change languages:
1.
Press
MENU key
to display a “MAIN MENU” screen.
If the
screen is not showing, use
Ø
Ø
or
×
×
key
to display it.
2.
Press
ENTER key
to display
.
3.
Press
Ø
Ø
key
until
screen appears.
4.
Press
ENTER key
to display
. Use
Ø
Ø
and
×
×
keys
to view the language choices.
5.
With the desired language displayed, press
ENTER
key
to enter this selection.
NOTE:
After a language is selected and entered, all
screens are displayed in that language.
The analyzer must be configured to define the sensor used
with it, and other related characteristics such as the tem-
perature element, desired buffer set for calibration, input
signal filtering, pulse suppression, etc.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.
ANALYZER CONFIGURATION