background image

Warning:

 Performance and reliability issues that result from the controller parameters being changed from 

the factory default settings are not covered under warranty.

Controller error codes:

P1- Room Probe Failure

Check to make sure the probe is connected to the controller 

Check the probe resistance (see “Checking the probe resistance” below)

P2- Evaporator Probe Failure

This error should only occur on MFV units. Verify that the unit is an MFV model.

Check to make sure the probe is connected to the controller 

Check the probe resistance (see “Checking the probe resistance” below)

dF- Indicates that the unit is in defrost

Normal Scheduled defrosts can last up to 20 minutes at the factory default values

If abnormal defrosts are suspected see “Resetting the program to factory default settings” below 

HA- Maximum Temperature Alarm

The box temperature has exceeded the 75° maximum for more than 30 minutes

See diagnostic flow charts below

LA- Minimum Temperature Alarm

The box temperature has deceeded the 25° minimum (-15° for MFV) for more than 30 minutes

See diagnostic flow charts below

Checking the controller set point:

Press the set button once to display the controller set point. Press again to return to the temp display.

Changing the controller set point:

Hold the SET key for 3 seconds, release, and use the arrow keys to change the set point. The set point can be 
changed from 33° to 65° on coolers and 0° to 28° on MFV's

Factory Default set points are as follows: MFV 10°, Beer 34°, White Wine 50°, Red Wine 60°

There is a 5° (8° for MFV) differential between cut in and cut out temps. If the controller set point is 35° 
the range will be 35° to 40°. (MFV controller set point 10°, range will be 10° to 18°)

Checking the probe resistance:

Verify the probe resistance is accurate at the box temperature

Use a calibrated thermometer to check the cooler's temperature at the evap inlet where the probe is located

Disconnect the probe from the controller 

Use a calibrated Ohm meter to measure the resistance of the probe at the box temperature

The resistance of the probe should match the associated box temperature of the following table

Fill a cup full of ice water (use a lot of ice), put the probe into the ice bath, stir for 1 minute, then measure the 
resistance with a calibrated Ohm meter. Make sure to keep the probe in the center of the cup.

The probe should have 27,280 Ohms of resistance in the ice bath. Replace the probe if it is not within 2 
degrees Ohms range from Table 1 above.

F-755-028

Rev. E

9/16/16

Table 1

Temp Deg F

Temp Deg F

Temp Deg F

Temp Deg F

Temp Deg F

-4

67770

22

34852

42

21596

64

13246

86

8313

-2

64579

23

33900

44

20687

66

12668

88

8008

0

61388

24

33164

46

19778

68

12090

90

7703

2

58197

26

31693

48

18869

70

11626

92

7398

5

53410

28

30222

50

17960

72

11161

94

7093

8

49763

30

28751

52

17233

74

10697

95

6940

10

47332

32 (Ice Water)

27280

54

16507

76

10232

96

6816

12

44901

34

26118

56

15780

77

10000

98

6569

14

42470

36

24956

58

15053

78

9813

100

6322

16

40566

38

23793

59

14690

80

9438

102

6074

18

38661

40

22631

60

14401

82

9063

104

5827

Ohms 

Resistance

Ohms 

Resistance

Ohms 

Resistance

Ohms 

Resistance

Ohms 

Resistance

Отзывы: