8. Troubleshooting and Maintenance Services

Please call the after-sales telephone and contact after-sales service personnel of 

MELING BIOLOGY & MEDICAL for replacement. Please do not replace the parts by 

yourself.

The following conditions are not faults

 When the compressor starts and stops, the equipment parts will make a slight impact sound;

 When overheated items are put in after the door is opened, high temperature alarm and high humidity alarm will 

be triggered in the control system (if this function is available, please refer to the alarm display table);

Solution: 

Put the items into the equipment after they are naturally cooled. Put the items in small quantities in 

batches, and do not put too much at a time. After the system runs stably, the high temperature alarm and high humidity 
alarm will be released automatically. 

 Slight sound of running water caused by refrigerant flowing in the pipeline.

¿

The equipment can only be repaired, maintained or improved by the engineers certified by MELING BIOLOGY & 
MEDICAL, so as to ensure the normal operation of the equipment and the compliance with corresponding safety 
standards.

¿

Please clean and disinfect the equipment before notifying the maintenance engineer; During the warranty period of 
the equipment, the Company will not undertake the warranty obligation if the fault or damage is caused by improper 
use of the user.

¿

Ambient temperature for storage: -40°C-+55°C, relative humidity: 10% ~ 90%.

Any product may fail. Please observe the operation of equipment in time during use. If there is any abnormality, please 

check and handle it according to the following table first. If the abnormality can't be changed, please inform our service 

center in time, and we will serve you wholeheartedly to avoid losses.

Fan motor replacement time: about 6 years; Service life of the product: 10 years.

9. Specifications

Notes:

20

21

After the equipment is used for a period of time, the door seal becomes hard and deformed.

Maintenance method: Blow the deformed part of the hot seal with a blower to soften it, and 

then close and compress it after the door seal becomes soft.

Is the power outlet charged?

Is the power plug plugged in or loose? 

Is the power fuse disconnected? 

Is the supply voltage too low or too high?

Is the temperature set correctly? 

Is the temperature inside the cabinet too low?

Are the doors not tightly closed or open too many times? 

Are there too many items put in at one time? 

Is the ambient temperature too high?

Is the cabinet placed on a flat ground? 

Does the cabinet contact the wall?

In rainy and humid seasons, door condensation is normal, and it shall be wiped off 

with a dry cloth.

Equipment needs to be cleaned.

Items with heavy smell are not packed.

Is the temperature set correctly? 

Is the temperature inside the cabinet too low?

Equipment does not 

work

Compressor is not 

running

The temperature 

does not reach the 

set value

High noise

Undesirable odor

The door is not 

closed properly, and 

the cool air leaks

Condensation on 

cabinet surface

The alarm lamp 

flashes and the 

buzzer is triggered

Lighting damage 

(refrigeration 

equipment)

Problems

Causes and solutions

* The foaming material of this product is cyclopentane.

Model 

Ambient 

temperature 

Climate 

type 

Refrigerant 

and loading 

amount 

Rated 

voltage 

(v~) 

Rated 

frequency 

(Hz) 

Temp 

Range 

(°C) 

Effective 

volume 

(L) 

Input 

power 

(W) 

Weight 

(Kg) 

Exterior 

dimensions 

(DxWxH) (mm) 

YC-55L 

16 ~32°C 

R600a/14g  220~230V 

50 

2~8 °C 

55 

121 

35 

560X540X632 

YC-55EL  16 ~32°C 

R600a/14g  220~230V 

50 

2~8 °C 

55 

125 

32 

560X540X632 

YC-75L 

16 ~32°C 

R600a/18g  220~230V 

50 

2~8 °C 

75 

147 

41 

560X540X764 

YC-75EL  16 ~32°C 

R600a/18g  220~230V 

50 

2~8 °C 

75 

135 

35 

560X540X764 

YC-130L  16 ~32°C 

R134a/110g  220~230V 

50 

2~8 °C 

130 

155 

51 

625X650X810 

YC-315L  16 ~32°C 

R600a/22g  220~230V 

50 

2~8 °C 

315 

285 

87 

673X650X1762 

YC-395L  16 ~32°C 

R600a/25g  220~230V 

50 

2~8 °C 

395 

290 

95 

673X650X1992 

YC-395EL  16 ~32°C 

R600a/25g  220~230V 

50 

2~8 °C 

395 

250 

89 

673X650X1992 

YC-400L  16 ~32°C 

R290/65g  220~230V 

50 

2~8 °C 

400 

500 

116 

645X700X2016 

YC-525L  16 ~32°C 

R290/75g  220~230V 

50 

2~8 °C 

525 

548 

148 

720X810X1961 

YC-725L  16 ~32°C 

R290/85g  220~230V 

50 

2~8 °C 

725 

790 

171  758X1093X1972 

YC-1015L  16 ~32°C 

R290/140g  220~230V 

50 

2~8 °C 

1015 

796 

223  900X 1180X 1990 

YC-1015EL  16 ~32°C 

R290/140g  220~230V 

50 

2~8 °C 

1015 

667 

206  900X 1180X 1990 

YC-1505L  16 ~32°C 

R507/500g  220~230V 

50 

2~8 °C 

1505 

1400 

322  830X1795X1990 

Содержание GL-RG-12M

Страница 1: ...GL RG 4M GL RG 12M GL RG 16M GL RG 23M OPERATION MANUAL Applicable product model www aegisfridge com ...

Страница 2: ...roduct composition and overview 6 Operating Instructions 6 1 Function Introduction 6 2 Optional function 7 Maintenance and Service 7 1 Equipment Maintenance 7 2 Equipment Discontinuation 7 3 Maintenance Replacement and Recovery of Rechargeable Batteries 8 Troubleshooting and Maintenance Services 9 Specifications 10 Packing List 11 Electrical Schematic Diagram 1 2 4 4 4 4 5 5 6 7 10 10 19 19 20 19 ...

Страница 3: ...er products For your safe and convenient use and reasonable maintenance of the equipment please read the Operation Instructions carefully before use and keep it properly for reference The equipment operator can copy some chapters of this operation manual but only for internal use for example to instruct the user how to deal with emergencies These chapters are clearly marked in the catalog of the m...

Страница 4: ...utlet tightly to ensure firmness and reliability so as to avoid electric leakage After installation the power plug must be within reach so that the power plug can be unplugged in time in case of emergency Separate special outlets must be used and grounded reliably The cross sectional area of the copper 2 conductor in the wall connected with the outlet must be more than 4mm Do not lengthen the powe...

Страница 5: ...r any outlet of the equipment otherwise electric shock or injury may be caused due to accidental contact between the above objects and moving parts In order to ensure the normal operation and ventilation and heat dissipation of the equipment the back left and right sides of the cabinet shall be at least 30cm away from the wall and the air inlet and air outlet must not be blocked by obstacles This ...

Страница 6: ... cold and heat sources and strong electromagnetic interference which will affect the normal operation of the control system and directly damage the system in severe cases 4 2 Installation Site Figure N and E ends cannot be connected In order to operate the equipment normally and obtain the best performance level the installation site of the equipment shall meet the following requirements ò It cann...

Страница 7: ...rt both ends of the handle cover into the corresponding grooves of the handle seat and then press it to close it 4 3 Preparation before use 4 4 First Power on When using the equipment for the first time please follow these steps 1 After the equipment is placed leveled and cleaned it shall stand for more than 24 hours and then power it on to ensure the normal operation of the equipment 2 Under no l...

Страница 8: ...reached Electrical appliances without production license shall not be used inside the equipment Do not change the set temperature frequently in a short time otherwise the expected setting effect may not be achieved due to the large temperature inertia Ensure that there is a certain air circulation space around the cabinet when putting in items especially do not block the temperature sensor in the ...

Страница 9: ...M Figure II GL RG 4M USB interface Door body Display panel Door lock Test hole Cabinet Caster Adjust feet Power switch Remote alarm interface Rs485 interface Door lock Test hole Door handle Cabinet Adjust feet Caster Door body USB interface Display panel ...

Страница 10: ...8 Figure III Figure IV Caster Door handle Door body Cabinet Caster Door lock Adjust feet Steel wire shelf Test hole Printer optional Display panel power switch on the back USB interface ...

Страница 11: ...he product consists of cabinet door glass door structure or foam door structure refrigeration system and control system Scope of application It is suitable for storing items in hospitals pharmacies epidemic prevention stations research institutions biopharmaceuticals and other units Figure V Steel wire shelf Door body Caster Cabinet Test hole Printer optional Display panel power switch on the back...

Страница 12: ...ff When the printer is working the indicator is on f Door heating indicator When the door heating is turned off the door heating indicator is off When the door heating is turned on the door heating indicator is on g Refrigeration indicator If the compressor is in working condition the refrigeration indicator is on If the compressor is in a shutdown state the cooling indicator is off h Mute indicat...

Страница 13: ...y for 3 seconds to import the data of the USB flash drive in 12 months 5 is a down key In parameter setting mode move to the previous parameter or decrease the parameter value For example when setting the set temperature reduce the set temperature When setting the parameter value long press the down button and the parameter will decrease rapidly 6 is a print key The system can keep 7 days of data f...

Страница 14: ...0 5 0 Temperature setting 2 H 0 0 10 0 5 0 Set value of high temperature alarm set H When H 0 High temp alarm is disabled When the alarm is over high temp alarm set H1 will be displayed on the controller 3 L 0 0 10 0 4 0 Set value of low temperature alarm set L When L 0 Low temp alarm is disabled When the alarm is below low temp alarm set L1 will be displayed on the controller 4 Pt 0 240 min 20 Pr...

Страница 15: ...arm L1 Low temperature alarm H2 Alarm for high ambient temperature H3 Condenser overheat alarm do Door opening alarm PF Power failure alarm bL Battery low alarm Er The recorder is not connected LoF Battery low alarm EE Communication failure 5 Alarm display ...

Страница 16: ...t above the key is off e PRINT optional The system can keep 7 days of data for printing Press the print key to print the temperature within the set time period f In parameter setting mode reduce the parameter value For example when setting the set temperature reduce the set temperature When setting the parameter value long press the down button and the parameter will decrease rapidly g In paramete...

Страница 17: ...d displays LoF the recorder is not started Meanwhile press and key for 3s and LoF disappears the recorder is started 2 Function setting of Type B control panel 1 Power on and turn on the power switch on the back of the cabinet and the equipment will enter the working state 2 User parameter settings Unlock In normal operation state simultaneously press and key for 3s and the digital tube displays t...

Страница 18: ...is opened and closed again during the heating period the heating time will be updated again Mode 2 When the compressor runs Heater is on When the compressor stops the heater Will automatically off after one min Mode 3 When the humidity in the cabinet is more than 80 the door heating is on and when the humidity in the cabinet is moderately less than 60 the door heating is off Mode 4 Door heating is...

Страница 19: ...ting paper has been installed at the factory When the paper is used up after a long time you may purchase the same paper for installation The installation steps are as follows Click the cylindrical button on the printer to open the cover of the printer Put the printing paper into the printer box and pull the paper head of the roll so that the paper slightly exposes from the upper port of the box c...

Страница 20: ...and RS485 interface The remote alarm terminal is installed on the back of the cabinet and the alarm signal is output by the terminal The load bearing capacity of the terminal is 30V 2A DC Users can choose normally open normally closed and common terminals according to their own needs 18 RS 485 ...

Страница 21: ...month and the power lock switch should be turned on so that the equipment can run for a period of time to complete charging and each charging time should be no less than 24 hours When the working power supply is interrupted the power lock switch of the equipment shall be turned off in time otherwise it will cause power loss in battery and even permanent damage to the battery in severe cases The ma...

Страница 22: ...roduct may fail Please observe the operation of equipment in time during use If there is any abnormality please check and handle it according to the following table first If the abnormality can t be changed please inform our service center in time and we will serve you wholeheartedly to avoid losses Fan motor replacement time about 6 years Service life of the product 10 years Notes 20 After the eq...

Страница 23: ...540X764 YC 130L 16 32 C N R134a 110g 220 230V 50 2 8 C 130 155 51 625X650X810 YC 315L 16 32 C N R600a 22g 220 230V 50 2 8 C 315 285 87 673X650X1762 YC 395L 16 32 C N R600a 25g 220 230V 50 2 8 C 395 290 95 673X650X1992 YC 395EL 16 32 C N R600a 25g 220 230V 50 2 8 C 395 250 89 673X650X1992 YC 400L 16 32 C N R290 65g 220 230V 50 2 8 C 400 500 116 645X700X2016 YC 525L 16 32 C N R290 75g 220 230V 50 2 ...

Страница 24: ...mity Accessory package Key The door handle accessory applies to models YC 725L YC 1015L YC 1015EL and YC 1505L Keys apply to models YC 55L YC 55EL YC 75L YC 75EL YC 130L YC 315L YC 395L YC 395EL YC 400L YC 525L YC 725L and YC 1505L The specific packing list shall be subject to the physical objects received 1 1 1 2 Operation Manual ...

Страница 25: ...EL YC 400L YC 525L YC 725L YC 1015L YC 1015EL YC 1505L Applicable to YC 130L The technical data and electrical schematic diagram shall be subject to the nameplate on the product and the electrical schematic diagram in case of any improvement of the product design We sincerely apologize for any inconvenience that might be caused ...

Отзывы: