Placement and installation
Right placement
The correct positioning of the balance greatly influences the
accuracy of the measurements.
It is important to ensure a stable and anti-vibration support.
To avoid
•
Direct light
•
Temperature fluctuations
•
Dust
•
Conditioners and fans
Antitheft
The balance is equipped with an anti-theft system, designed to
discourage a possible attempted theft.
The cable and the padlock are not supplied with the scale.
Levelling the balance
The BL 224 has a level control and adjustable feet to compensate for
minor irregularities in the weighing surface.
The scale is exactly horizontal when the air bubble is in the middle of
the target.
Before turning on the balance, level it by rotating the adjustable feet so
that the level air bubble is in the center of the ring.
Make sure all the feet are securely fixed on the bench.
Connection of peripheral electronic devices
Turn off the balance
before connecting or disconnecting a device
(printer, PC).
Before performing the first measurement, the weighing sensor
temperature must align with the ambient temperature in order to
obtain the best results.
Warm up:
60 min.