Automatic Calibration
Automatic calibration can works based on:
•
Time
Automatic calibration after a set time;
•
Temperature
Automatic calibration when the temperature changes;
•
Calibration when power on
Automatic calibration every power on
Auto Cal Time setting
•
Press the <
ON / OFF
> key, and during the ignition press the <
CAL
> key
•
The display shows "
A.CAL OFF
", use the
▲
key to move to "
A.CAL On
", and
press <
TARE
> to confirm.
•
"
TIME
" appears, press <
TARE
> to enter.
•
• The <
UNIT
> key scrolls through the possible values, with <
TARE
> selects
the desired one:
Time selection:
•
0,5 h (30 minutes)
•
1 h
•
2 h
•
3 h
•
24 h
•
Off (function deactivated)
Auto Cal Temperature settings
Following the same procedure as above, select "
A.CAL On
", and using the
▲
key
select "
Temp
".
Temperature selections:
•
0,5 °C
•
1 °C
•
2 °C
•
5 °C
•
Off (function deactivated)
Power on AUTO Cal
Following the same procedure as above, when "
A.CAL On
" or "
A.CAL Off
" appears,
press the <CAL> key to enter the calibration settings at power up.
When "
P.Cal On
" is selected, the balance will be calibrated automatically at each
power-up.