INTRODUCTION & S U G G E S T I O N S
If items are damaged or if you have questions about the operation or maintenance of the unit
STOP and contact us immediately!
This operation manual provides necessary information for the safe use of the
Gio Peel
Microdermabrader however, it is
meant to supplement, not replace a formal training program in the use of this device.
General Description
The Gio Peel Microdermabrasion system is designed to exfoliate by abrading the skin with high-purity grade Aluminum Oxide
Crystal or may also be used for crystal-free operation. The major components of the system include:
• Powerful diaphragm vacuum pump
• One simple vacuum flow control valve
• Plastic crystal volume flow control screw nut (no adjusting needed after initial adjustment)
• Prefilled and sealed “Crystal Capsule
™
” containing enough abrading media for 7 to 9 complete treatments
• Crystal Capsule
™
also contains a sealed used crystal waste receptacle located on the bottom half of the disposable capsule
• Disposable plastic handpiece with connecting tubing
• Disposable Crystal-Free Kit
Not only is your Gio Peel lightweight and easy to move from room to room, the handpiece is very easy to manipulate and its size and
shape has been engineered to provide precision and accuracy for the microdermabrasion procedure. The tip is designed to provide
the proper angle to abrade the epidermis safely and gradually. The tubing for the handpiece is connected to two posts located in the
center front housing of the Crystal Capsule. The tubing connects the system’s crystal supply to the handpiece then is vacuumed up
and deposited in the waste section of the same capsule! Only when the tip is in contact with the skin is the vacuum created by clos
-
ing the loop after connecting the disposable Crystal Capsule the vacuum port post on the front of the Gio Peel. The vacuum gauge
indicates the vacuum level that is regulated by the flow control valve.
How To Get The Most From This Manual
Properly follow the instructions in the order listed below, to avoid injuries or jeopardize any warranty coverage.
• Use this manual as a check off list. Check mark each task when completed.
• Concentrate on the material, pre-read the manual first before doing anything else.
• Read in an isolated area away from distractions.
• Do it our way first. We’ve spent a lot of time developing the most efficient methods for you.
• Be prepared. Have everything ready before starting.
• Take notes to make the subject clearer or to enhance the material.
• Learn at your own pace. Don’t give up. It will come to you as practice makes perfect.
• Learn from your mistakes. Correct the error and try again.
• Don’t take chances. Only perform a step when it is clear exactly what to do.
Hints, Tips & Precautions
• Always wear a N95/N99 NIOSH approved respirator, gloves and ANSI Z87.1 approved eye protection.
• The media recommended for use with this device is aluminum free.
• Only operate unit in temperature ranges of 60°F to 85°F.
• Never, under any circumstance, operate with wet or damp hands or while standing in liquid.
• Only use a 3 wire grounded electrical outlet capable of delivering 115vac at 10 amps.
• Never operate the Gio Peel while plugged into a 2 wire extension cord.
•
NEVER LOAD OR UNLOAD THE GIO PEEL WHILE THE UNIT IS OPERATING.
• Only operate this unit with dry media recommended by Gio Pelle, Inc.
• Unplug the Gio Peel from the wall electrical outlet when not in use.
• Inspect the equipment for damage and wear each time before using.
• Verify the Gio Peel media size is properly suited for the patient or application.
• Never operate the Gio Peel unattended.
• Only use according to the instructions given in this manual.
• Never operate with liquid media or liquid cleaners.
• Do not attempt to alter or modify the unit or operate it with any sort of attachments or devices unless specifically provided or
recommended by the manufacturer.
• Do not operate this device with a damaged cord or plug or after the unit malfunctions or has been damaged. If damaged, return
the unit for examination, repair or adjustment.