PERIODICAL MAINTENANCE CHECKS
The
TOBI MANUALE
suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use.
Unpack the instrument and
always check
integrity of plastic parts, they might have been damaged
during previous use.
Close the aspirator outlet with your finger.
Press the foot and check vacuum functionally.
Type of fault
Cause
Remedy
1. No aspiration
Jar Cap badly screwed down
Unscrewed the cap, then
rescrew it correctly
2. No aspiration
Lid seal not in its seat
Unscrew the cap and insert the
seal properly in its seat
3. The float doesn’t close
If the cap has been washed,
ensure that the float is not
partially detached
Insert the float into it’s place
4. The float doesn’t close
The float it’s covered by dirty
material
Unscrewed the cap, leave the
and put in on autoclave
5. Low suction
Foam inside the jar
Fill the jar to 1/3 full of ordinary
water
6. No aspiration due to
flow leakage of mucus
Filter blocked
Replace filter
Faults 1 - 2 - 3 - 4 - 5 - 6
None of the remedies has
achieved the desired results
Contact the seller or GIMA
After-sales Assistance
Service
If the overfill security system it’s activated, don’t proceed with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter
who avoid the liquid penetration inside the device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device,
in this case return the device to GIMA S.p.A. technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA S.p.A. TECHNICAL SERVICE.
GIMA S.p.A. DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.