MO003-GB 18/05/06
Operative Manual
Rev.0
Page 26 of 34
Dimensions
Lamp body diameter [cm]
23
Internal reflector diameter [cm]
16
Light emission surface [cm
2
] 200
Maximum head vertical movement [cm]
146
Scialytic lamp weight [Kg]
22,5
Certificates
In conformity with the 93/42/CEE
directive
Chapter 5 – Maintenance
Ordinary maintenance of the apparatus consists of:
•
changing the halogen bulb when burnt-out (Paragraph 4.3)
•
changing the fuses in case of breakage (Paragraph 4.4)
•
calibrating the oscillating arm (Paragraph 5.1)
5.1 Calibrating the oscillating and horizontal arm
The lamp is sold already balanced and no further calibration is
necessary. If over time the oscillating arm with spring balance becomes
rigid or loose it is possible to intervene mechanically by adjusting the
compression of the internal spring.
Bring the oscillating arm to the highest position allowed and remove
aluminium cover (1).
Insert a pin of 5mm maximum and rotate it anti-clockwise or clockwise
to increase or diminish the strength of the spring.
At the end of adjustment replace the cover in the original position.
After adjustment the balancing should be fluid.
1
Содержание SATURNO WALL
Страница 30: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 30 of 34 6 2 Explode drawing...
Страница 31: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 31 of 34...
Страница 32: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 32 of 34...
Страница 33: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 33 of 34...
Страница 34: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 34 of 34...