MO003-GB 18/05/06
Operative Manual
Rev.0
Page 20 of 34
Checking installation:
1 Check the suitability of the wall on installing the lamp
2 The wall must guarantee a load of 10 Kg to 200cm distance
3 Check that the arrangement of the wall anchorage is suitable for fixing the lamp
4 The lamp must be fixed to the wall with 4 nuts and 4 counter-nuts at each point. Every
counter-nut must be fixed with a drop of strong thread delay Loctite model 270.
5 Check the horizontally of the plate in relation to the wall, using a level
6 Verify the correct graft of screw TSCE M3x15 on the horizontal arm (in the lateral part) in the
center of the electrical contact so as to avoid of the twisting of cables
7 Check the removal of the stop inserted in the oscillating arm once installation is completed
8 Check the presence of a protection (magneto-thermal) on the primary and the fuses on the
secondary as instructed (at the client’s expense)
9 Check the supply tension on the transformer output terminals and verify that is between 12,0
and 13,0 Volt
10 Check the correct insertion of the halogen bulb (flush on the lamp-holder and on the vertical)
11 Take care not to touch the halogen bulb with your fingers. Use a dry cloth to insert the
halogen bulb into the bulb-holder
15 Check the light emission glass of the head of the lamp
In the case in which the light emission glass is damaged or broken, fragments of glass might
fall into the patient zone. The lamp must be turned off immediately and no longer used until
the fault has been eliminated.
16 Check that the mechanical movement element functions perfectly.
Check the mechanical functioning of the lamp with orientation and rotation movements of the
mechanical movement element.
Scialytic test:
1 After turning on, the lamp should supply light
Tension read at contacts (12,0<V
1.2
<13,0):
2 Positioning the head at 1m from the protective glass of
the luminous intensity probe and identifying the
maximum luminous intensity zone (centre of the
luminous band), position the luminous field at
maximum intensity by rotating the handle that can be
sterilised.
klx reading (I = 40 klx ± 5%):
Stamp and signature of the installer
__________________________________
Содержание SATURNO WALL
Страница 30: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 30 of 34 6 2 Explode drawing...
Страница 31: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 31 of 34...
Страница 32: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 32 of 34...
Страница 33: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 33 of 34...
Страница 34: ...MO003 GB 18 05 06 Operative Manual Rev 0 Page 34 of 34...