16
SpO2, heart rate
SpO2 BPM
---- % -----
OPERATIONAL SCREEN
Display Messages guide
CHECK PADS
The defibrillation pads are disconnected.
IMPEDANCE
Indicates the impedance out of range (min 20 ohm, max 200 ohm).
Verify the pads contact with the patient chest and add more conductive
gel.
ARMING
RESCUE 230 is charging the capacitor to the energy level selected.
READY
Indicates that the capacitor is charged and the device is ready for
defibrillation. In Sync mode the shock will be released only when a valid
ECG wave R peak is detected. In this mode make sure a valid ECG trace
is available on the display.
BAT OK/LOW
Indicates the battery status. If
LOW
then please connect the AC power
supply.
DATE
TIME
HEART
RATE
ECG TRACE
ENERGY SELECTION
BATT STATUS AND FILTERS
SPEED AND GAIN OF THE
ECG TRACE