Gima PRIMALED Скачать руководство пользователя страница 15

 

 

Operative Manual 

MO079-IT 

15/06/12 

Rev.0 

Pag. 15 di 16 

 

 

 

10.

 

Friction regulation 

The  device  is  provided  already  balanced  and  no  further  adjustments  are  required;  however,  if  over  time  the 
movement  of  arms  around  their  rotation  joints  should  became  too  rigid  or  too  soft,  so  as  to  not  allow  the 
maintenance  in  position  of  the  device  itself,  it  is  possible  to  act  on  the  different  friction  systems  in  order  to 
restore the correct balance. 

 

Use the included Allen key to adjust the friction force in correspondence of rotational joints and, therefore, the 
consequent movement of movable arms.  

 

ROTATION JOINTS 

 

Depending on the version of the device, there is a different number of joints and so of frictions: 

 

    

    

            

 

 
 
 
 

FRICTION ADJUSTING PROCEDURE 

 

 

In  correspondence  of  the  joint  of  interest, 
using  the  Allen  key,  act  on  the  regulating 
screw on the side of joint. 
Rotate  clockwise  to  increase  the  force  of 
friction  and  hence  stiffen  the  movement. 
Rotate  counter-clockwise  to  decrease  the 
friction force and make lighter the movement. 
After  the  adjustment  the  movement  must  be 
fluid and homogeneous. 

 
 

 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 

Table version 

Table version with flexible arm  

Содержание PRIMALED

Страница 1: ...ation mobile version 6 2 2 Installation wall version S 12MED fixing 6 2 3 Installation wall version rail bar fixing 6 2 4 Installation table version S11 fixing 6 2 5 First starting 6 2 6 Installation...

Страница 2: ...alian edition of this operator s manual 1 General information The ME Medical Electrical EQUIPMENT to which this manual refers is a LAMP for diagnosis or observation For easier description such ME EQUI...

Страница 3: ...tion manual and user manual Transport is carried out by any road haulage contractor as long as they respect the following characteristics Temperature C 15 60 Humidity 10 75 Atmospheric pressure h Pa 5...

Страница 4: ...ential performance Collateral standard Electromagnetic compatibility Requierements and tests Classification with reference to article 9 and Annex IX of Directives 93 42 EEC and 2007 47 EC DURATION Sho...

Страница 5: ...ot above 40 of the net value of the product as indicated on the buyer s invoice GIMA s liability is expressly ruled out for indirect damages or consequential damages including cases of the lamp not be...

Страница 6: ...the bar and tighten the handle 3 Insert the lamp into the upper hole of the rail clamp 4 Follow points 2 and 3 of paragraph 2 1 2 4 Installation table version S11 fixing 1 Fix the clamp S 11 to the ta...

Страница 7: ...ions could cause glare and potential damage to the retina Never place and or hang anything on the Product Unless this precaution is taken positioning will not be reliable and the danger exists of such...

Страница 8: ...ic parts becoming fragile use only disinfectants with low alcohol content Dose the disinfectants so no liquids penetrate inside the lamp bodies and into the support arm system Clean the Product with a...

Страница 9: ...tch on Check the INPUT and OUTPUT 24V power supply voltage and the voltage of the electronic board terminal 24V 2 The Product does not remain in position See the attached instructions 3 The light flic...

Страница 10: ...240 Frequency Hz 50 60 Absorbed power VA 30 Light source N 12 Led x 1 4W LED light source duration h this datum can vary according to voltage peaks and the frequency of use 50 000 Light intensity con...

Страница 11: ...Operative Manual MO079 IT 15 06 12 Rev 0 Pag 11 di 16 8 Electrical diagram...

Страница 12: ...ngs used for domestic purposes provided the following warning is heeded Warning This equipment system is intended for use by healthcare professionals only This equipment system may cause radio interfe...

Страница 13: ...cial or hospital environment NOTE UT is the a c main voltage prior to application of the test level Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted R...

Страница 14: ...transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 24 0 1 0 38 0 38 0 73 1 1 2 1...

Страница 15: ...to adjust the friction force in correspondence of rotational joints and therefore the consequent movement of movable arms ROTATION JOINTS Depending on the version of the device there is a different n...

Страница 16: ...Operative Manual MO079 IT 15 06 12 Rev 0 Pag 16 di 16 Note...