Operative Manual
MO079-IT
15/06/12
Rev.0
Pag. 12 di 16
9.
EMC declaration
The Product has been tested in accordance to EN60601-1-2 to ensure proper electromagnetic compatibility.
Portable and mobile RF-communications equipment can affect the Product. Other products used in the vicinity of
Product should also comply with this standard.
The Product is intended for use in the electromagnetic environment specified below. The customer or the user of
the Product should assure that these are used in such an environment.
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The Product uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The Product is suitable for use in all establishments
other than domestic, and may be used in domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes,
provided the following warning is heeded:
Warning:
This equipment/system is intended for
use by healthcare professionals only. This
equipment/system may cause radio interference or
may disrupt the operation of nearby equipment. It
may be necessary to take mitigation measures, such
as re-orienting or relocating the Product or shielding
the location
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Immunity test
IEC 60601-1-2
test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete
or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least 30%
Electrical fast transient /
burst IEC 61000-4-4
+/- 2 kV
for power supply unit
+/- 1 kV
for input/output lines
+/- 2 kV
for power supply
lines
+/- 1 kV
for input/output lines
Mains power quality should be
that of a typical commercial or
residential environment.
Surge
IEC 61000-4-5
+/- 1 kV
differential mode
+/- 2 kV
common mode
+/- 1 kV
differential mode
+/- 2 kV
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.