5
• This monitor is not suitable for use during the operation of MRI or CT.
This Monitor is not MRI Compatible.
• To ensure that the monitor is safely grounded when charging use
a three-pole socket.
• Only use manufacturer designated accessories to ensure compliance
with appropriate standards
• It is not allowed to remove the cover of the monitor.
• This monitor provides End tidal CO
2
(EtCO
2
) concentration, Respiration
Rate, Oxygen saturation and Pulse Rate. This data only provides assistance
for diagnosis and actual diagnosis shall be made by suitably qualified
clinical staff using all the clinical information and symptoms.
• In order to prevent pressure sores and correct circulation the SpO
2
sensor
must be repositioned regularly, depending on the type of sensor used.
Chapter II
Technical specifications and Characteristics
EtCO
2
Method:
Creative proprietary non-dispersive InfraRed
Spectroscopy
Range:
0 – 150mmHg or 0 – 20kPa or 0 – 20% (v/v)
Accuracy:
±2mmHg for EtCO
2
range 0 - 40mmHg
±5% for EtCO
2
range from 41 - 70mmHg
±8% for EtCO
2
range from 71 - 100mmHg
Over 100mmHg ±10%
Note
: The accuracy of CO
2
concentration measurement is influenced by
any interfering gas and/or vapour, for example N
2
O gas can raise the CO
2
reading (2-10%), and Helium and O
2
can reduce the CO
2
reading (1-10%),
so compensation should be set in the balance gas MENU to meet the
accuracy requirements if such gases or vapours are present.
Update/Averaging Time: Option of every breath or 10, 20 or 30 seconds
Warm Up Time:
<20 seconds
Sample Flow Rate:
50–250ml/min User Adjustable. Default =100ml/min
Patient Modes:
Adult and Neonatal
Memory:
24 hours on Screen Trend and Numeric
Sensor:
<25g Single Use Gas Sample Line and Adapter for
Intubated and /or Non Intubated Patients
Color Change:
Connector housing has additional built-in LED
Colour change feature. Purple (or Red) during
occlusion or pump err.
Respiration Rate
Range:
3 - 150 breaths/minute