25
This product is not suitable for mechanical re-processing or sterilization.
Cleaning instructions:
Use Moist (not dripping) wipes with 70% solution of isopropyl alcohol, or very
dilute Chlor-clean (1000ppm) or Chlorohexidine (1000ppm) or mild detergent,
then allow to air dry naturally.
3. Cleaning of filter/water trap T3 if required (only for re-use)
Please Note: It is advised to use the filter/water trap T3 as a single
patient use item so as to absolutely remove the risk of cross infection.
Re-Use Cleaning:
Only the filter surface may be cleaned and/or disinfected. Use Moist
(not dripping) wipes with 70% solution of isopropyl alcohol, or very dilute
Chlor-clean (1000ppm) or Chlorohexidine (1000ppm) or mild detergent,
then allow to air dry naturally.
Chapter VII
Trouble Shooting Analysis
Simple analysis of problems
No. Phenomena
Causes
Solution
1
2
3
4
5
The values of CO2 is reading
too low, or 'OCCLUSION'
appears on the screen.
The values of CO2 is zero.
1. Screen indicating PUMP
ERR and big noise.
2. Screen indicating
IR-LAMP-BAD.
3. Screen indicating CO2
SENSOR ERR.
Screen indicating CAL-ERR.
Screen indicating
POWER-ERR.
The CO2 wave is not normal.
1. Screen indicating TEMP-
HIGH.
1. Leaking of filter or sampling
tube.
2. Occlusion of filter
or sampling line.
3. Out of Calibration.
1. Internal leaking inside
the Pump gas loop.
2. The IR lamp resource
of sensor damaged.
3. IR Sensor broken.
The last calibration has failed.
Damaged or incorrect power
supply.
1. Temperature too high.
2. Temperature too low.
3. Sharp ambient Temperature
change.
1. Check and replace filter
T3 or sample line.
2. Clear the gas loop
Occlusion.
3. Re-calibrate using
standard gas.
Contact the Distributor
or manufacturer for repair.
Re-calibrate using standard
gas.
Contact Distributor
or manufacturer.
Use in normal environmental
temperature range.