It is recommended to carry out the treatments after medical consultation.
A simultaneous connection of a patient to an high frequency surgical unit can
produce burnings in correspondence of the electrodes and the unit can be
damaged.
This equipment was manufactured to satisfy the present electromagnetic
compatibility. In case of the functioning of the unit is interfered or interfy with
the normal functioning of other electrical and electronic equipments, it is
recommended to feed the equipment by a different electrical plug and/or try to
dispose the unit in a different way until to the interference ends.
Do not use mobile phone near to the equipment to avoid possible interference
with it.
Many components used inside this equipment are sensible to the electrostatic
charges. If you need to hand any electronic component, avoid the eventual
static energy that you can carry, by touching a mass plane to charge at hearth.
If it is possible, wear a bracelet connected to earth. If you do not observe these
precautions you could cause permanent damages to the electronic components
of the unit.
Working near a short wave therapy or microwaves unit (for example: lower
than 2 m) can produce instability in output.
Before every use check the unit and cables condition to determine discovered
conductor parts or other condition of electric risk.
The unit does not fit explosive or gas environments.