GIMA SpA
IONO BASE PLUS
36 Manuale
d’Istruzione/Instruction’s
Manual
MA342IGBa
D
ISPLAY
C
HANNEL
CH1
(3)
On top at the right of the unit there is a display to the visualization of the intensity, in
mA. At the switching on of the unit the intensity is at 0 mA and can be changed (from 0
to 20, by single step) by the suitable keys, when the unit does not supply. The + key is
used to increase the intensity value, while the – key to decrease it.
D
ISPLAY
C
HANNEL
CH2
(4)
In the central part of the unit, at the right, there is a display to the visualization of the
intensity, in mA. At the switching on of the unit the intensity is at 0 mA and can be
changed (from 0 to 20, by single step) by the suitable keys, when the unit does not
supply. The + key is used to increase the intensity value, while the – key to decrease it.
D
ISPLAY
T
IME
T
IMER
(5)
At the left of the unit, there is a display to visualize the time values, in minutes.
Switching on the unit, the time is at 30 minutes and can be changed (from 0 to 60 min)
by the suitable keys, when the apparatus does not supply. The + key is used to increase
the frequency value, while the – to decrease it. By the start key
(5c)
it is possible to
start the time count and start the output too. The output start condition is indicated by
the lighting of a light point at right of minutes’ display. By the stop key
(5d),
it is
possible to stop the time count and stop the output too. The stop condition is indicated
by the switching off of the point at the right of the minutes’ display. The end of the
treatment time is indicated by an acoustic signal of 10 seconds. By setting the 00 time
it’s possible to realize an infinite treatment time, without programmed interruption.
C
ONNECTORS
OUT
1
(6)
Points of connections for the electrodes’ cables. The red connectors must be associated
to the positive polarity and the black to the negative one. The warning light near the
output 1 connectors indicates the current supply and it is lighted when the circuit is
closed (connected patient).
C
ONNECTORS
OUT
2
(7)
Points of connections for the electrodes’ cables. The red connectors must be associated
to the positive polarity and the black to the negative one. The warning light near the
output 2 connectors indicates the current supply and it is lighted when the circuit is
closed (connected patient).